BN-129 Abel Automatic Closed Flash Point Tester
whose products are widely used in petrochemical industry, national defense science and technology, quality inspection and customs, scientific research institutions, third-party testing and other industries.
keyword:
Automatic distillation _ instrument _ automatic distillation, trace saturated vapor pressure
The classification:
Home show
BN-129 Abel Automatic Closed Flash Point Tester
Applicable standards:
GB/T 6536 GB/T 3146 GB/T 7534 ASTM D86 ASTM D850 ASTM D1078
Product description:
BN-129 Abel Automatic Closed Flash Point Tester complies with IP 170 and EN ISO13736 standards, accurately determines the flash point from -30 ° C to 110 ° C, with high degree of automation, and can be widely applied to petroleum products, alcohols, solvents, chemicals. , food and other tests.
Structural features:
● This instrument is a desktop structure as shown.
●Using 8.4-inch LCD screen, all Chinese display.
● It adopts industrial grade mainboard control, which integrates temperature control, ignition, flash point detection, storage and printing as one and completes automatically. Automatic detection of flash point: automatic opening and ignition.
●Each instrument is equipped with two ignition systems, free to choose electronic or gas igniter; if gas igniter is used, the electronic igniter will automatically re-ignite the gas flame as needed during the test.
● External flame sensors increase the safety of the system.
●With multi-function circulating cooling water bath, automatic / manual control of cooling medium circulation.
● The test results are automatically corrected to the standard atmospheric pressure; after the test is completed, the test results are automatically saved, and the results can be uploaded to the LIMS system.
Technical Parameters:
◆Test range: -30 ° C ~ 110 ° C (external circulation cooling water bath)
◆ Temperature resolution: 0.1 ° C
◆Ignition: gas ignition and electric ignition (interchangeable)
◆ Heating: according to method or user customization
◆ Stirring speed: according to the method or user-defined
◆ Flash point detection: thermocouple differential detection;
◆Temperature measurement: PT100
◆Atmospheric pressure correction: Automatically correct the whole machine power: 300W
◆ Machine power: 300W
◆Safety: automatic flame detection, automatic fire extinguishing
◆Working power supply: AC220V±10%, 50Hz
◆Dimensions: 530 × 392 × 286 (mm)